				Fluke Calibration 9190A-D-156
                
                
                
 			 
				Catalog #: 4178901
DRY-WELL, ULTRA-COLD, INSRT-D, 115V
 
 
                
                    OverviewSpecificationsSummarySoftwareCasesInsertsAccessories 	
                    
				        
                        Product Highlights 9190A 
    Temperature range: -95 °C to      140 °C
Built-in reference      thermometer readout accuracy: ± 0.05 °C full range
Display accuracy: ± 0.2 °C      full range
Stability: ± 0.015 °C full      range
Cooling time: 23 °C to -90      °C: 80 minutes; 23 °C to -95 °C: 90 minutes; 140 °C to 23 °C: 60 minutes
Portability: Weighs 16 kg (35      lb)
Built-in front and back      handles for easy two-handed carry
Conforms to EURAMET cg-13      guidance on measurement practices for temperature block calibrators
All controls and inputs on      front panel
Stability indicator shows      whether block temperature is stable and within limits defined by the user      PC interface
USB and RS-232 support remote      operation from computer
Active heater zone control      compensates for differential temperatures between zones
Minimizes axial temperature      gradients
Define up to eight set-point      temperatures; automatically ramp to and "soak" at each
Specify soak time, number of      cycles, and direction
Fast time to temperature and      small footprint
Compact, efficient free      piston Stirling cooler technology (FPSC)
 9190-INSD 
    Six metric comparison holes
Hole quantities and sizes: 2      Holes, 3 mm.; 2 Holes, 4 mm.; 2 Holes, 6 mm
Removable inserts sized for      probes with named diameter
Use additional inserts to      broaden range of hole sizes available
For use with Model 9190A Each Unit Includes 
    1 - "D" Insert1 - Power Cord1 - USB Cable1 - User Manual CD1 - 9930 Interface-it Software1 - Report of Calibration1 - Well Insulator1 - Insert Removal Tool4 - Clamp Ferrites 
                        
    
        
            | Base unit specifications |  
            | Temperature range at 23 °C | –95 °C to 140 °C (–139 °F to 284 °F) |  
            | Display  accuracy | ± 0.2 °C full range |  
            | Accuracy with  external reference [3] | ± 0.05 °C full range |  
            | Stability | ± 0.015 °C full range |  
            | Axial uniformity at 40 mm (1.6  in) | ± 0.05 °C full range |  
            | Radial gradient | ±0.01 °C full range |  
            | Loading effect | (with a 6.35  mm reference probe  and three  6.35  mm probes) |  
            | ± 0.006 °C full range |  
            | (versus display  with one 6.35  mm probe) |  
            | ± 0.25 °C at –95 °C |  
            | ± 0.10 °C at 140 °C |  
            | Operating conditions | 0 °C to 35 °C, 0 % to 90 % |  
            | RH (non-condensing) < 2000 m altitude |  
            | Environmental conditions for all specifications except temperature range | 13 °C to 33 °C |  
            | Immersion (well)  depth | 160  mm (6.3 in) |  
            | Well diameter | 30 mm (1.18 in) |  
            | Heating time [1] | –95 °C to 140 °C: 40 min |  
            | Cooling  time [1] | 23 °C to –90 °C: 80 min |  
            | 23 °C to –95 °C: 90 min |  
            | 140 °C to 23 °C: 60 min |  
            | Stabilization time [2] | 15 min |  
            | Resolution | 0.01  ° |  
            | Display | LCD, °C or °F user selectable |  
            | Size (H x W x D) | 480  mm x 205  mm x 380  mm (18.8 in x 8.0 in x 14.9  in) |  
            | Weight | 16 kg (35 lb) |  
            | Power requirements | 100 V to 115 V (± 10 %) 50/60 Hz, 575 W |  
            | 200 V to 230 V (± 10 %) 50/60 Hz, 575 W |  
            | System fuse  ratings | 115 V: 6.3 A T 250 V |  
            | 230 V: 3.15  A T 250 V |  
            | 4–20 mA Fuse  (-P model only) | 50 mA F 250 V |  
            | Computer interface | RS-232, USB Serial, and 9930 interface-it temperature calibration software included |  
            | Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |  
            | Electromagnetic environment | IEC 61326-1: Basic |  
            | Refrigerants | R32 (Difluoromethane) |  
            | < 20 g, ASHRAE safety group A2L |  
            | R704 (Helium) |  
            | < 20 g, ASHRAE safety group A1 |    
    
        
            | -P Specifications |  
            | Built-in reference thermometer readout accuracy (4-wire reference probe) [3] | ± 0.010 °C at –95 °C |  
            | ± 0.013 °C at –25 °C |  
            | ± 0.015 °C at 0 °C |  
            | ± 0.020 °C at 50 °C |  
            | ± 0.025 °C at 140 °C |  
            | Reference resistance range | 0 Ω to 400 Ω |  
            | Reference resistance accuracy [4] | 0 Ω to 42 Ω: ± 0.0025 Ω 42 Ω to 400  Ω: ± 60 ppm of reading
 |  
            | Reference characterizations | ITS-90, CVD, IEC-751, resistance |  
            | Reference measurement capability | 4 wire |  
            | Reference probe connection | 6-pin din with INFO-CON technology |  
            | Built-in RTD thermometer readout accuracy | NI-120: ± 0.015 °C at 0 °C |  
            | PT-100 (385): ± 0.02 °C at 0 °C |  
            | PT-100 (3926):  ± 0.02 °C at 0 °C |  
            | PT-100 (JIS): ± 0.02 °C at 0 °C |  
            | RTD resistance range | 0 Ω to 400 Ω |  
            | Resistance accuracy [4] | 0 Ω to 25 Ω: ± 0.002 Ω |  
            | 25 Ω to 400 Ω: ± 80 ppm of reading |  
            | RTD characterizations | PT-100 (385), (JIS), (3926), NI-120, resistance |  
            | RTD measurement capability | 2-wire, 3-wire, and 4-wire RTD with jumpers  only |  
            | RTD connection | 4-terminal input |  
            | Built-in TC thermometer readout accuracy [5] | Type J: ± 0.70 °C at 140 °C |  
            | Type K: ± 0.75 °C at 140 °C |  
            | Type T: ± 0.60 °C at 140 °C |  
            | Type E: ± 0.60 °C at 140 °C |  
            | Type R: ± 1.60  °C at 140  °C |  
            | Type S: ± 1.60  °C at 140  °C |  
            | Type M: ± 0.65  °C at 140  °C |  
            | Type L: ± 0.65  °C at 140  °C |  
            | Type U: ± 0.70 °C at 140 °C |  
            | Type N: ± 0.75 °C at 140 °C |  
            | Type C: ± 1.00 °C at 140 °C |  
            | TC millivolt range | –10  mV to 75 mV |  
            | Voltage accuracy | 0.025 % of reading +0.01 mV |  
            | Internal cold  junction compensation accuracy | ± 0.35 °C (ambient of 13 °C to 33 °C) |  
            | TC connection | Miniature connectors (ASTM E1684) |  
            | Built-in mA readout accuracy | 0.02  % of reading + 0.002 mA |  
            | mA range | Cal 4-22 mA, Spec 4-24 mA |  
            | mA connection | 2 terminal  input |  
            | Loop power function | 24 V dc loop power |  
            | Built-in electronics temperature coefficient (0 °C to 13 °C, 33 °C to 50 °C)
 | ± 0.005 % of range  per °C |  Notes:[1]   For ambient temperature of 23 °C.
 [2]   Time from when the SETPOINT is reached to when the unit is with in stability  specification.
 [3]   The temperature range  may be limited by the reference probe  connected to the readout.
 The built-in reference accuracy  does not include  the sensor  probe  accuracy.  It does not include  the probe  uncertainty or probe  characterization errors.
 [4]   Measurement accuracy  specifications apply within  the operating range  and assume 4 wires  for PRTs. With 3-wire RTDs add 0.05  Ωto the measurement accuracy  plus the maximum possible  difference between the resistances of the lead wires.
 [5]   The thermocouple input readout is sensitive to EM fields in the frequency range  of 500  MHz to 700  MHz.
   
    
        
         
            | Calibration Included | Factory Calibration |  
            | Product Weight | 38 LBS |  
            | ECCN Number | EAR99 |  
            | Country of Origin | United States |  |  
					
                
        
        
              Software
              Cases
              Inserts
              Accessories 
				 |  
     |  |