				Fluke Calibration 9190A-F-P-156
                
                
                
 			 
				Catalog #: 4179045
DRY-WELL,ULTRA-COLD,INSRT-F,BLT-IN ELCT, 115V
 
 
                
                    OverviewSpecificationsSummarySoftwareCasesInsertsAccessories 	
                    
				        
                        Product Highlights9190A 
    Temperature range: -95 °C to      140 °CBuilt-in reference      thermometer readout accuracy: ± 0.05 °C full rangeDisplay accuracy: ± 0.2 °C      full rangeStability: ± 0.015 °C full      rangeCooling time: 23 °C to -90      °C: 80 minutes; 23 °C to -95 °C: 90 minutes; 140 °C to 23 °C: 60 minutesPortability: Weighs 16 kg (35      lb)Built-in front and back      handles for easy two-handed carryConforms to EURAMET cg-13      guidance on measurement practices for temperature block calibratorsAll controls and inputs on      front panelStability indicator shows      whether block temperature is stable and within limits defined by the user      PC interfaceUSB and RS-232 support remote      operation from computerActive heater zone control      compensates for differential temperatures between zonesMinimizes axial temperature      gradientsDefine up to eight set-point      temperatures; automatically ramp to and "soak" at eachSpecify soak time, number of      cycles, and directionFast time to temperature and      small footprintCompact, efficient free      piston Stirling cooler technology (FPSC) 9190-INSF 
    Six metric/imperial      comparison miscellaneous holesHole quantities and sizes: 2      Holes, 3 mm; 2 Holes, 4 mm; 1 Hole, 6 mm; 1/4 in.Removable inserts sized for      probes with named diameterUse additional inserts to      broaden range of hole sizes availableFor use with Model 9190A Process Electronics            
    Reference sensor control:
    
        Set-point temperature       control gets passed from internal sensor to external reference PRT9190A controls well       temperature based on its measurement and position inside wellHelps minimize effects of       axial gradient when reference PRT is aligned with short sensors4-20 mA connectors:Power 4-20 mA transmitter       from drywellSupplies 24 V dc loop power4-wire PRT/RTD input:
    
        Reads 4-wire, 3-wire, or       2-wire PRT/RTDAccuracy ± 0.02 °C4-20 mA circuit fusesThermocouple input:
    
        Accepts common thermocouples       with mini-jack terminationTypes: J, K, T, E, R, S, M,       L, U, N, CReference thermometer input:Smart connector input       accepts -A terminated reference PRTsAccuracy ± 0.01 °C at -95 °C Each Unit Includes
    1 - "F" Insert1 - Power Cord1 - USB Cable1 - User Manual CD1 - 9930 Interface-it Software1 - Report of Calibration1 - Well Insulator1 - Insert Removal Tool4 - Clamp Ferrites1 - 6-PIN DIN Connector1 - Test Lead Kit 
                        
    
        
            | Base unit specifications |  
            | Temperature range at 23 °C | –95 °C to 140 °C (–139 °F to 284 °F) |  
            | Display  accuracy | ± 0.2 °C full range |  
            | Accuracy with  external reference [3] | ± 0.05 °C full range |  
            | Stability | ± 0.015 °C full range |  
            | Axial uniformity at 40 mm (1.6  in) | ± 0.05 °C full range |  
            | Radial gradient | ±0.01 °C full range |  
            | Loading effect | (with a 6.35  mm reference probe  and three  6.35  mm probes) |  
            | ± 0.006 °C full range |  
            | (versus display  with one 6.35  mm probe) |  
            | ± 0.25 °C at –95 °C |  
            | ± 0.10 °C at 140 °C |  
            | Operating conditions | 0 °C to 35 °C, 0 % to 90 % |  
            | RH (non-condensing) < 2000 m altitude |  
            | Environmental conditions for all specifications except temperature range | 13 °C to 33 °C |  
            | Immersion (well)  depth | 160  mm (6.3 in) |  
            | Well diameter | 30 mm (1.18 in) |  
            | Heating time [1] | –95 °C to 140 °C: 40 min |  
            | Cooling  time [1] | 23 °C to –90 °C: 80 min |  
            | 23 °C to –95 °C: 90 min |  
            | 140 °C to 23 °C: 60 min |  
            | Stabilization time [2] | 15 min |  
            | Resolution | 0.01  ° |  
            | Display | LCD, °C or °F user selectable |  
            | Size (H x W x D) | 480  mm x 205  mm x 380  mm (18.8 in x 8.0 in x 14.9  in) |  
            | Weight | 16 kg (35 lb) |  
            | Power requirements | 100 V to 115 V (± 10 %) 50/60 Hz, 575 W |  
            | 200 V to 230 V (± 10 %) 50/60 Hz, 575 W |  
            | System fuse  ratings | 115 V: 6.3 A T 250 V |  
            | 230 V: 3.15  A T 250 V |  
            | 4–20 mA Fuse  (-P model only) | 50 mA F 250 V |  
            | Computer interface | RS-232, USB Serial, and 9930 interface-it temperature calibration software included |  
            | Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |  
            | Electromagnetic environment | IEC 61326-1: Basic |  
            | Refrigerants | R32 (Difluoromethane) |  
            | < 20 g, ASHRAE safety group A2L |  
            | R704 (Helium) |  
            | < 20 g, ASHRAE safety group A1 |    
    
        
            | -P Specifications |  
            | Built-in reference thermometer readout accuracy (4-wire reference probe) [3] | ± 0.010 °C at –95 °C |  
            | ± 0.013 °C at –25 °C |  
            | ± 0.015 °C at 0 °C |  
            | ± 0.020 °C at 50 °C |  
            | ± 0.025 °C at 140 °C |  
            | Reference resistance range | 0 Ω to 400 Ω |  
            | Reference resistance accuracy [4] | 0 Ω to 42 Ω: ± 0.0025 Ω 42 Ω to 400  Ω: ± 60 ppm of reading
 |  
            | Reference characterizations | ITS-90, CVD, IEC-751, resistance |  
            | Reference measurement capability | 4 wire |  
            | Reference probe connection | 6-pin din with INFO-CON technology |  
            | Built-in RTD thermometer readout accuracy | NI-120: ± 0.015 °C at 0 °C |  
            | PT-100 (385): ± 0.02 °C at 0 °C |  
            | PT-100 (3926):  ± 0.02 °C at 0 °C |  
            | PT-100 (JIS): ± 0.02 °C at 0 °C |  
            | RTD resistance range | 0 Ω to 400 Ω |  
            | Resistance accuracy [4] | 0 Ω to 25 Ω: ± 0.002 Ω |  
            | 25 Ω to 400 Ω: ± 80 ppm of reading |  
            | RTD characterizations | PT-100 (385), (JIS), (3926), NI-120, resistance |  
            | RTD measurement capability | 2-wire, 3-wire, and 4-wire RTD with jumpers  only |  
            | RTD connection | 4-terminal input |  
            | Built-in TC thermometer readout accuracy [5] | Type J: ± 0.70 °C at 140 °C |  
            | Type K: ± 0.75 °C at 140 °C |  
            | Type T: ± 0.60 °C at 140 °C |  
            | Type E: ± 0.60 °C at 140 °C |  
            | Type R: ± 1.60  °C at 140  °C |  
            | Type S: ± 1.60  °C at 140  °C |  
            | Type M: ± 0.65  °C at 140  °C |  
            | Type L: ± 0.65  °C at 140  °C |  
            | Type U: ± 0.70 °C at 140 °C |  
            | Type N: ± 0.75 °C at 140 °C |  
            | Type C: ± 1.00 °C at 140 °C |  
            | TC millivolt range | –10  mV to 75 mV |  
            | Voltage accuracy | 0.025 % of reading +0.01 mV |  
            | Internal cold  junction compensation accuracy | ± 0.35 °C (ambient of 13 °C to 33 °C) |  
            | TC connection | Miniature connectors (ASTM E1684) |  
            | Built-in mA readout accuracy | 0.02  % of reading + 0.002 mA |  
            | mA range | Cal 4-22 mA, Spec 4-24 mA |  
            | mA connection | 2 terminal  input |  
            | Loop power function | 24 V dc loop power |  
            | Built-in electronics temperature coefficient (0 °C to 13 °C, 33 °C to 50 °C)
 | ± 0.005 % of range  per °C |  Notes:[1]   For ambient temperature of 23 °C.
 [2]   Time from when the SETPOINT is reached to when the unit is with in stability  specification.
 [3]   The temperature range  may be limited by the reference probe  connected to the readout.
 The built-in reference accuracy  does not include  the sensor  probe  accuracy.  It does not include  the probe  uncertainty or probe  characterization errors.
 [4]   Measurement accuracy  specifications apply within  the operating range  and assume 4 wires  for PRTs. With 3-wire RTDs add 0.05  Ωto the measurement accuracy  plus the maximum possible  difference between the resistances of the lead wires.
 [5]   The thermocouple input readout is sensitive to EM fields in the frequency range  of 500  MHz to 700  MHz.
   
    
        
         
            | Calibration Included | Factory Calibration |  
            | Product Weight | 38 LBS |  
            | ECCN Number | EAR99 |  
            | Country of Origin | United States |  |  
					
                
        
        