				Additel ADT878PC-425-D-110V
                
                
                
 			 
				Catalog #: ADT878PC-425-D-110V
Reference Dry Well Calibrator
 
 
                
                    OverviewSpecificationsSummary 	
                    
				        
                        Product Highlights
    Temperature ranges:
    
        160: -40 to 320°F425: 91.4 to 797°F700: 91.4 to 1292°FAccuracy:
    
        160: ±0.1°C at full range425: ±0.2°C at Full Range700: ±0.20°C at 33°C/425°C and ±0.25°C at 660°CFull HART communicatorWi-Fi and Bluetooth capableColor touch screen displayQuick-Push connectors (PC Option)Set point control by referenceSelf calibrationBuilt-in automatic annealing (ADT878-700 only) Each Unit Includes
    Dry block and selected insertPower cableUSB cableInsert removal toolCertificate of calibrationCD manual 
                          
    
        
            | Temperature Range at 23°C | 33°C to 425°C |  
            | Display Accuracy | ±0.2°C at Full Range |  
            | Stability (30 min) | ±0.005°C at 100°C ±0.010°C at 225°C
 ±0.015°C at 425°C
 |  
            | Axial Uniformity at 60 mm (2.4 in) | ±0.10°C at 100°C ±0.15°C at 225°C
 ±0.25°C at 425°C
 |  
            | Axial Uniformity at 80 mm (3.15 in) | ±0.15°C at 100°C ±0.20°C at 225°C
 ±0.30°C at 425°C
 |  
            | Radial Uniformity | ±0.025°C at 100°C ±0.030°C at 225°C
 ±0.040°C at 425°C
 |  
            | Lading Effect | ±0.05°C (Display Sensor) ±0.01°C (External Sensor
 |  
            | Hysteresis (Display Sensor) | 0.04°C |  
            | Environmental Conditions | 8°C to 38°C guaranteed accuracy 0°C to 50°C, 0% to 90% RH non-condensing
 |  
            | Storage Conditions | -20°C to 60°C |  
            | IP Rating | IP20 |  
            | Immersion Depth | 193 mm (7.60 in) |  
            | Insert OD | 30.8 mm (1.21 in) |  
            | Heating Time | 15 min: 23°C to 425°C |  
            | Cooling Time | 24 min: 425°C to 100°C 15 min: 100°C to 50°C
 |  
            | Typical Time to Stability | 10 min |  
            | Resolution | 0.001°C |  
            | Units | °C, °F, and K |  
            | Display | 6.5 in (165 mm) color touch screen |  
            | Size (H x W x D) | 170 x 345 x 330 mm (6.69 x 13.58 x 13.0 in) |  
            | Weight | 9.7 kg (21.4 lbs) |  
            | Power Requirements | 90-254 VAC, 45-65 Hz, 1400 W |  
            | Communication | USB A, USB B, RJ45, WiFi, Bluetooth |  
            | Localization | English, Chinese, Japanese, Russian, German, French, Italian, and Spanish |  
            | Warranty | 1 yr |    
    
        
            | Input Specifications (Process Calibrator [PC] Option) |  
            | Readout Accuracy for 100 ohm PRT (Probe Accuracy Not Included) | ±0.006°C at 0°C ±0.008°C at 50°C
 ±0.009°C at 100°C
 ±0.011°C at 160°C
 ±0.015°C at 300°C
 ±0.019°C at 425°C
 ±0.026°C at 660°C
 ±0.028°C at 700°C
 |  
            | Readout Resolution | 0.1 mΩ |  
            | Reference Resistance Range | -200Ω to 926Ω |  
            | Reference Resistance Accuracy | 0Ω to 50Ω: ±1.25mΩ 50Ω to 400Ω: ±0.0025% RD
 |  
            | Reference Characterizations | ITS-90, CVD, IEC-751 |  
            | Reference Measurement Capability | 4-wire PRT |  
            | Reference Probe Connection | 6-pin lemo smart connector and Quick-Push connectors to accept banana, mini-banana, large & small spade lug and bare wire connections |  
            | RTD Channels | 2 |  
            | RTD Measurement Accuracy (excl sensor) Compliance | 0Ω to 50Ω: ±2.5mΩ 50Ω to 4000Ω: ±0.005% RD
 |  
            | RTD Measurement Resolution | 0.1mΩ |  
            | RTD Measurement Resistance Range | 0Ω to 4KΩ |  
            | RTD Characterizations | PT10, PT25, PT50, PT100, PT200, PT500, PT1000, CU10, CU50, CU100, NI100, NI120 |  
            | RTD Connection | Quick-Push connectors accept banana, mini-banana, large & small spade lug and bare wire connections |  
            | RTD Channels | 2 channels. Both accept 2, 3, or 4-wire RTDs |  
            | TC Channel | 2 |  
            | TC Measurement Channels | Accepting S, R, K, B, N, E, J, T, C, D, G, L, and U |  
            | TC Range | -75 mV to 75 mV |  
            | TC Resolution | 0.1µV |  
            | TC Voltage Accuracy | 0.01% RD + 5 µV |  
            | Internal CJC Accuracy | ±0.2°C (ambient from 0°C to 50°C) |  
            | Current Range | -30 mA to 30 mA |  
            | Current Accuracy | 0.01% RD + 2 µA |  
            | Current Resolution | 0.1 µA, Input Impedance: < 10Ω |  
            | Voltage Range | -30 V to 30 V |  
            | Voltage Accuracy | ±0.01% RD + 0.6 mV |  
            | Voltage Resolution | 0.1 mV; Input impedance: >1MΩ |  
            | Switch Test | Mechanical or Electrica |  
            | DC 24V Output | 24 V ±0.5 V, MAX 60 mA |  
            | Hart Communicator | Read, configure and calibrate HART devices - DD files updated periodically Optional - (order ADT875PC) |  
            | Documentation | Up to 1,000 tasks which store up to 10 results each containing as found and as left data. Snap shot feature allows for screen captures. Records auto step and ramp functions. |  
            | Temperature Coefficient 0℃ to 13℃ and 33℃ to 50℃ | ADT878 (PC)-160: ±0.005°C/°C ADT878 (PC)-425/700: ±0.005°C/°C
 Ref Readout: ±1 ppm FS/°C
 RTD Readouts: ±1 ppm FS/°C
 TC Readouts: ±5 ppm FS/°C
 Current: ±5 ppm FS/°C
 Voltage: ±5 ppm FS/°C
 |  
    
        
            | Input Specifications (Process Calibrator [PC] Option) |  
            | TC Type | Temperature (°C) | Error (°C) [1] |  
            | B | 250 | ±1.99 |  
            | 300 | ±1.65 |  
            | 425 | ±1.18 |  
            | 660 | ±0.81 |  
            | 700 | ±0.77 |  
            | 1768 | ±0.56 |  
            | K | -200 | ±0.29 |  
            | -40 | ±0.13 |  
            | 0 | ±0.13 |  
            | 160 | ±0.14 |  
            | 300 | ±0.15 |  
            | 425 | ±0.16 |  
            | 660 | ±0.18 |  
            | 700 | ±0.19 |  
            | 1000 | ±0.31 |  
            | E | -200 | ±0.16 |  
            | -40 | ±0.09 |  
            | 0 | ±0.09 |  
            | 160 | ±0.08 |  
            | 300 | ±0.09 |  
            | 425 | ±0.10 |  
            | 660 | ±0.12 |  
            | 700 | ±0.13 |  
            | 1000 | ±0.17 |  
            | J | -210 | ±0.22 |  
            | -40 | ±0.10 |  
            | 0 | ±0.10 |  
            | 160 | ±0.11 |  
            | 300 | ±0.12 |  
            | 425 | ±0.13 |  
            | 660 | ±0.14 |  
            | 700 | ±0.14 |  
            | 1000 | ±0.21 |  
            | T | -200 | ±0.28 |  
            | -40 | ±0.14 |  
            | 0 | ±0.13 |  
            | 160 | ±0.11 |  
            | 300 | ±0.11 |  
            | 400 | ±0.11 |  
            | N | -200 | ±0.46 |  
            | -40 | ±0.20 |  
            | 0 | ±0.19 |  
            | 160 | ±0.17 |  
            | 300 | ±0.17 |  
            | 425 | ±0.17 |  
            | 660 | ±0.19 |  
            | 700 | ±0.19 |  
            | 1000 | ±0.27 |  
            | S | -50 | ±1.25 |  
            | -40 | ±1.17 |  
            | 0 | ±0.93 |  
            | 160 | ±0.63 |  
            | 300 | ±0.57 |  
            | 425 | ±0.55 |  
            | 660 | ±0.54 |  
            | 700 | ±0.53 |  
            | 1768 | ±0.66 |  
            | R | -50 | ±1.33 |  
            | -40 | ±1.23 |  
            | 0 | ±0.95 |  
            | 160 | ±0.61 |  
            | 300 | ±0.54 |  
            | 425 | ±0.51 |  
            | 660 | ±0.48 |  
            | 750 | ±0.18 |  
            | 1768 | ±0.58 |  
            | [1] Excluding cold junction compensation errors. |    
    
        
            | HTS/Schedule B Number | 9025.80.10 |  |  
					
				 
				 |  
     |  |