Display: 3? digit liquid crystal display (LCD) with a maximum reading of 1999.
Polarity: Automatic, positive implied, negative polarity indication.
Overrange: (OL) or (-OL) is displayed.
Zero: Automatic.
Low battery indication: The B is displayed when the battery voltage drops below the operating level.
Measurement rate: 2.5 per second, nominal.
Operating environment: 0 ?C to 50 ?C at 70 % R.H.
Storage temperature: -20 ?C to 60 ?C, 0 to 80 % R.H. with battery removed from meter.
Temperature Coefficient: 0.1 ? (specified accuracy) per ?C. (0 ?C to 18 ?C, 28 ?C to 50 ?C).
Environment: Indoor use, Altitude up to 2000 m
Power: Single standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22.
Battery life: 200 hours typical with carbon-zinc.
Dimensions: 196 x 92 x 60 mm ( 7.7? x 3.5? x 2.4?).
Weight: Approx. 426 .( 0.94 lb.) without holster, including battery.
Warranty: One (1) Year
DC Voltage Ranges: 200 mV, 2 V, 20 V, 200 V, 600 V
DC Voltage Accuracy: all ranges: ? (1 % rdg + 1 dgt)
DC Voltage Resolution: 100 µV
DC Voltage Input Impedance: 10 M?
AC Voltage (45 Hz to 500 Hz) Ranges: 200 mV, 2 V, 20 V, 200 V, 600 V
AC Voltage (45 Hz to 500 Hz) Accuracy: all ranges: ? (1.5 % rdg + 4 dgts)
AC Voltage (45 Hz to 500 Hz) Resolution: 100 µV
AC Voltage (45 Hz to 500 Hz) Input Impedance: 10 M?
DC Current Ranges: 200 µA, 2 mA, 20 mA, 200 mA, 10 A
DC Current Accuracy: 200 µA to 200 mA ranges: ? (1.5 % rdg + 1 dgt)
DC Current Accuracy: 10 A range: ? (2.0 % rdg + 3 dgts)
DC Current Resolution: 0.1 µA
DC Current Voltage Burden (Max): 200 µA range: 1 mV/µA
DC Current Voltage Burden (Max): 2 mA range: 100 mV/1mA
DC Current Voltage Burden (Max): 20 mA range: 13 mV/1 mA
DC Current Voltage Burden (Max): 200 mA range: 4.6 mV/1 mA
DC Current Voltage Burden (Max): 10A range: 40 mV/1 A
AC Current (45 Hz to 500 Hz) Ranges: 200 µA, 2 mA, 20 mA, 200 mA, 10 A
AC Current (45 Hz to 500 Hz) Accuracy: 200 µA to 200 mA ranges: ? (2.0 % rdg + 4 dgts)
AC Current (45 Hz to 500 Hz) Accuracy: 10A range: ? (2.5 % rdg + 4 dgts)
AC Current Resolution: 0.1 µA
AC Current Voltage Burden: See DC Current
Non-Contact Voltage Detector (NCV) AC Volts: 70 to 600 V ac
Non-Contact Voltage Detector (NCV): Red LED and Audible Indicator
AC/DC Voltage: 400 mV range: 600 V dc or 600 V ac rms ]5 sec
All other ranges: 600 V dc or 600 V ac rms
AC/DC Current: uA/mA input: F 0.25 A/ 600 V Min. I.R. 30 kA (6.3 x 32 mm)
10 A input F 10A / 600V, Min. I.R. 100 kA, (10 x 38 mm)
(10 A for 4 minutes maximum followed by a 12 minute cooling period)
Resistance: 600 V dc or 600 V rms
Continuity: 600 V dc or 600 V ac rms
Diode Test: 600 V dc or 600 V ac rms
Battery Test: 600 V dc or 600 V ac rms
|