Display: 3 ¾ digit liquid crystal display (LCD)(3999 count) with a 41-segment analog bar-graph.
Polarity: Automatic, positive implied, negative polarity indication. Overrange = (0o) or (-0o) is displayed.
Zero: Automatic.
Low battery indication: The B is displayed when the battery voltage drops below the operating level.
Auto power off: Approx. 30 minutes.
Measurement rate: 2 times per second, nominal.
Operating environment: 0 º to 50 º at 70 % R.H.
Storage temperature: -20 º to 60 º, 0 to 80 % R.H. with battery removed from meter.
Temperature Coefficient: 0.1 x (specified accuracy) per º. (0 º to 18 º, 28 º to 50 º).
Altitude: 2000 m (6562 feet)
Power: Single standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22.
Battery life: 100 hours typical with carbon-zinc. 200 hours typical with alkaline. Frequent use of the backlight will decrease battery life.
Dimensions: 196 mm (H) x 92 mm (W) x 60 mm (D).
Weight: Approximately 400 g including battery.
Warranty: One-Year.
DC Voltage Ranges: 400 mV, 4 V, 40 V, 400 V, 1000 V
DC Voltage Accuracy: all ranges: ± (0.5 % rdg + 1 dgt)
DC Voltage Resolution: 100 µV
AC Voltage (45 Hz to 2 kHz) True RMS Ranges: 400 mV, 4 V, 40 V, 400 V, 750 V
AC Voltage (45 Hz to 2 kHz) True RMS Accuracy: 400 mV range, 45 to 100 Hz: ± (1.2 % rdg + 8 dgts)
AC Voltage (45 Hz to 2 kHz) True RMS Accuracy: 750 V range: 45 to 1 kHz: ± (2.0 % rdg + 8 dgts)
AC Voltage (45 Hz to 2 kHz) True RMS Accuracy: All other ranges, 45 to 500 Hz: ± (1.2 % rdg + 8 dgts)
AC Voltage (45 Hz to 2 kHz) True RMS Accuracy: All other ranges, 500 to 2 kHz: ± (2.0 % rdg + 8 dgts)
AC Voltage (45 Hz to 2 kHz) True RMS Resolution: 100 µV
DC Current Ranges: 400 µA, 4 mA, 40 mA, 300 mA, 10 A
DC Current Accuracy: 400 µA to 300 mA ranges: ± (1.0 % rdg + 1 dgt)
DC Current Accuracy: 10 A range: ± (2.0 % rdg + 3 dgts)
DC Current Resolution: 0.1 µA
DC Current Voltage Burden: 400 µA range: 1 mV/µA, 4 mA range: 100 mV/1 mA, 40 mA range: 12 mV/1 mA,
DC Current Voltage Burden: 400 µA, 4 mA, 40 mA, 300 mA, 10 A
AC Current (45 Hz to 1 kHz) True RMS Ranges: 400 µA, 4 mA, 40 mA, 300 mA, 10 A
AC Current (45 Hz to 1 kHz) True RMS Accuracy: 400 µA to 300 mA ranges: ± (1.5 % rdg + 8 dgts)
AC Current (45 Hz to 1 kHz) True RMS Accuracy: 10 A range: ± (2.5 % rdg + 10 dgts)
AC Current (45 Hz to 1 kHz) True RMS Resolution: 0.1 µA
Resistance Ranges: 400 ? 4 k? 40 k? 400 k? 4 M? 40 M?
Resistance Accuracy: 400, 4 k, 40 k, 400 k, 4 M? ranges: ± (1.0 % rdg + 4 dgts)
Resistance Accuracy: 40 M? range: ± (2.0 % rdg + 5 dgts)
Resistance Resolution: 100 m?
Continuity Audible Indication: 35 ?
Ranges: -20 °C to 1000 °C, -4 °F to 1832 °F
Accuracy: -20 °C to 10 °C: ± (2.0 % rdg + 4 °C )
10 °C to 200 °C: ± (1.0 % rdg + 3 °C )
200 °C to 1000 °C: ± (3.0 % rdg + 2 °C )
-4 °F to 50 °F: ± (2.0 % rdg + 8 °F)
50 °F to 400 °F: ± (1.0 % rdg + 6 °F)
400 °F to 1832 °F: ± (3.0 % rdg + 4 °F)
Resolution: 1 °C, 1 °F
Ranges: 4 kHz, 40 kHz, 400 kHz, 4 MHz, 40MHz
Accuracy: all ranges: ± (0.1 % rdg + 3 dgts)
Resolution: 1 Hz
Ranges: 0 to 90 %
Accuracy: 5 V logic: ± (2.0 % rdg + 5 dgts)
Frequency: Range 40 Hz to 20 kHz
Audible Indication: 35 ?
Test Current: 1.2 mA (approximate)
Open Circuit: Volts 3.0 V dc typical
AC/DC Voltage: All ranges 1000 V dc or 750 V ac rms
AC/DC Current: µA/mA input: 0.315 A /1000 V fast-blow ceramic (6.3 x 32 mm)
10 A input: 10 A /1000 V fast-blow ceramic 10 x 38 mm
(10 A for 4 minutes maximum followed by a 12 minute cooling period)
|