| Specifications | EX360 | EX363 | EX365 | 
        
            | Basic Accuracy (DCV) | 0.5% | 0.5% | 0.5% | 
        
            | NCV Detector | 80 to 1000V (50-60Hz) and 160 to 1000V (50-60Hz)
 | 80 to 1000V (50-60Hz) and 160 to 1000V (50-60Hz)
 | 80 to 1000V (50-60Hz) and 160 to 1000V (50-60Hz)
 | 
        
            | DC/AC Voltage | 0.1mV to 1000V | 0.1mV to 1000V | 0.1mV to 1000V | 
        
            | DC/AC Current | — | — | 1mA to 10A | 
        
            | DC/AC µA Current | — | 600µA | — | 
        
            | Resistance | 0.1Ω to 40MΩ | 0.1Ω to 40MΩ | 0.1Ω to 40MΩ | 
        
            | Capacitance | 1nF to 10mF | 1nF to 10mF | 1nF to 10mF | 
        
            | Frequency | 0.01Hz to 100kHz | 0.01Hz to 100kHz | 0.01Hz to 100kHz | 
        
            | Temperature (Type K) | — | -40 to 742°F (-40 to 394°C) | — | 
        
            | Diode Test | 1.5V | 1.5V | 1.5V | 
        
            | Continuity | Audible and Visual | Audible and Visual | Audible and Visual | 
        
            | Dimensions | 6.5x3.2x1.7" (164x82X44mm) | 6.5x3.2x1.7" (164x82X44mm) | 6.5x3.2x1.7" (164x82X44mm) | 
        
            | Weight | 5.9oz (450g) | 5.9oz (450g) | 5.9oz (450g) | 
        
            | ELECTRICAL SPECIFICATIONS | 
        
            | Accuracy is given as ± (% of reading + least significant digit) at 23 °C ±5 °C with relative humidity lower than 80%. Accuracy is specified for a period of one year after calibration.
 1. Temperature Coefficient is 0.1 x specified accuracy /  °C, < 18 °C (64.5 °F), > 28 °C (82.4 °F)
 2. AC Functionality: ACV and ACA specifications are AC coupled, True RMS; the accuracy for
 square waveforms is unspecified. For non-sinusoidal waveforms, additional accuracy Crest
 Factor (C.F.) considerations exist as detailed below:
 Add 3.0% for C.F. 1.0~2.0; Add 5.0% for C.F. 2.0~2.5; Add 7.0% for C.F. 2.5~3.0
 | 
        
            | Function | Range | Resolution | Accuracy (of reading) | ‘OL’ Reading | 
        
            | mV DC | 600.0mV | 0.1mV | ± (0.5% + 5 digits) | 660.0mV | 
        
            | Input Impedance: 10M; Overload Protection: AC/DC 1000V | 
        
            | DC Voltage | 6.000V | 0.001V | ± (0.5% + 2 digits) | 6.600V | 
        
            | 60.00V | 0.01V | 66.00V | 
        
            | 600.0V | 0.1V | 660.0V | 
        
            | 1000V | 1V | 1100V | 
        
            | Input Impedance: 10M; Overload Protection: AC/DC 1000V | 
        
            | AC Voltage | 600.0mV | 0.1mV | ± (1.0% + 5 digits) | 660.0mV | 
        
            | 6.000V | 0.001V | ± (1.0% + 3 digits) | 6.600V | 
        
            | 60.00V | 0.01V | 66.00V | 
        
            | 600.0V | 0.1V | 660.0V | 
        
            | 1000V | 1V | 1100V | 
        
            | LCD Displays ‘0’ counts when the reading < 10 counts Input Impedance: 10MΩ (< 100pF)
 Frequency Response: 45~500Hz (sine wave)
 Over load protection: 1000V AC/DC
 | 
        
            | Lo Z (Low Impedance) | 600.0V | 0.1V | ± (2.0% + 3 digits) | 660.0V | 
        
            | 1000V | 1V | 1100V | 
        
            | LCD Displays ‘0’ counts when the reading < 10 counts Input Impedance: < 3kΩ
 Frequency Response: 45~500Hz (sine wave); Over load protection: 1000V AC/DC
 | 
        
            | µA DC (EX363 only) | 600.0 µA | 0.1uA | ± (1.0% + 2 digits) | 660.0µA | 
        
            | Input Impedance: 3k approx. Over load protection: 1000V AC/DC
 | 
        
            | µA AC (EX363 only) | 600.0 µA | 0.1uA | ± (1.5% + 3 digits) | 660.0µA | 
        
            | LCD Displays ‘0’ counts when the reading < 10 counts Input Impedance: 3kΩ approx.
 Frequency Response: 45~500Hz (sine wave)
 Over load protection: 1000V AC/DC
 | 
        
            | AC Current (EX365) | 6.000 A | 0.001 | ± (1.5% + 3 digits) | 6.600A | 
        
            | 10.00 A | 0.01 | 20.00A | 
        
            | In 6A Range, the LCD displays 0 counts when the reading < 20 counts. In10A Range, the LCD displays 0 counts when the reading < 10 counts.
 Maximum measurement time: >5A for max. 3 minutes with at least 20 minutes rest time.
 >10A for max. 30 seconds with at least 10 minutes rest time.
 Frequency Response : 45 ~ 500Hz (Sine Wave)
 Overload Protection : AC/DC 11A
 | 
        
            | DC Current (EX365) | 6.000 A | 0.001 | ± (1.0% + 3 digits) | 6.600A | 
        
            | 10.00 A | 0.01 | 20.00A | 
        
            | Maximum measurement time : >5A for max. 3 minutes with at least 20 minutes rest time. >10A for max. 30 seconds with at least 10 minutes rest time.
 Overload Protection : AC/DC 11A
 | 
        
            | Resistance | 600.0 Ω | 0.1 Ω | ± (0.9% + 5 digits) | 660.0 Ω | 
        
            | 6.000k Ω | 0.001k Ω | ± (0.9% + 2 digits) | 6.600 kΩ | 
        
            | 60.00k Ω | 0.01k Ω | ± (0.9% + 2 digits) | 66.00 kΩ | 
        
            | 600.0k Ω | 0.1k Ω | ± (0.9% + 2 digits) | 660.0 kΩ | 
        
            | 6.000M Ω | 0.001M Ω | ± (0.9% + 2 digits) | 6.600 MΩ | 
        
            | 40.00M Ω | 0.01M Ω | ± (1.5% + 5 digits) | 44.00 MΩ | 
        
            | Overload Protection: 1000V AC/DC Note that digits may fluctuate ±50 digits when measuring above 10.00M
 | 
        
            | Continuity | 600.0 Ω | 0.1 Ω | ± (0.9% + 5 digits) | 660.0 Ω | 
        
            | Continuity: Built-in beeper sounds when measured resistance is less than 20O. Beeper turns off when measured resistance goes above 200O
 Continuity beeper frequency tone: 2KHz; Response Time: < 500µsec.; Overload Protection: AC/DC 1000V
 | 
        
            | Diode | 1.550V | 0.001V | ± (0.9% + 2 digits) | n/a | 
        
            | Open Circuit Voltage : Approx. 1.8V; Overload Protection : AC/DC 1000V | 
        
            | Capacitance | 1.000 µF | 0.001 µF | ± (1.9% + 5 digits) | 1.100 µF | 
        
            | 10.00 µF | 0.01 µF | ± (1.9% + 2 digits) | 11.00 µF | 
        
            | 100.0 µF | 0.1 µF | 110.0 µF | 
        
            | 1.000 mF | 0.001 mF | 1.100 mF | 
        
            | 10.00 mF | 0.01 mF | 11.00 mF | 
        
            | Overload Protection: 1000V AC/DC | 
        
            | Frequency | 100.00 Hz | 0.01 Hz | ± (0.1% + 2 digits) | 100.00 Hz | 
        
            | 1000.0 Hz | 0.1 Hz | 1000.0 Hz | 
        
            | 10.000 kHz | 0.001kHz | 10.000 kHz | 
        
            | 100.00 kHz | 0.01kHz | 100.00 kHz | 
        
            | Minimum Sensitivity: > 5V (for ACV 1Hz ~ 10kHz) > 20.0V (for ACV 10kHz ~ 50kHz) unspecified (for ACV 50kHz ~ 100kHz) >0.6A (for ACA)
 Minimum Frequency: 1Hz
 Overload Protection: AC/DC 1000V or 11A
 | 
        
            | Temp. (EX363) | -40.0~400.0  °C | 0.1° | ± (1.0% + 20 digits)* | 440.0  °C | 
        
            | -40.0~752.0  °F | 0.1° | ± (1.0% + 36 digits)* | 824.0  °F | 
        
            | *Does not include accuracy of the temperature probe. Accuracy specifications assume surrounding temperature stable to ±1 o C. For ambient temperature
 changes of ±2 o C, rated accuracy applies after 2 hours of stabilization time.
 
 OVERLOAD PROTECTION: AC/DC 1000V
 | 
        
            | Peak MIN-MAX |  | Specified accuracy ± 150 digits
 |  | 
        
            | Accuracy of square wave is unspecified | 
        
            | Non-Contact Voltage Detector
 | 80~1000V (high sensitivity mode) 50~60Hz | 
        
            | 160~1000V (low sensitivity mode) 50~60Hz | 
        
            | The tip of the meter offers the optimum sensitivity |