| General Specifications |
| |
OX 9062 |
OX 9102 |
OX 9104 |
OX 9304 |
| HUMAN-MACHINE INTERFACE |
| Type of display |
7‘’ WVGA colour TFT LCD touch screen, 800x480 – LED backlighting (adjustable standby mode) |
| Different display mode |
2,500 real acquisition points on screen - Vectors with interpolation |
| Display of curves on screen |
4 curves + 4 references – Split Screen & Full Screen modes |
| Screen commands |
Touch screen – ANDROID-type icons and graphical commands – customizable channel colours |
| Choice of language |
15 complete languages, menus & online help |
| OSCILLOSCOPE MODE |
| Vertical deflection |
| Bandwidth |
60 MHz |
100 MHz |
100 MHz |
300 MHz |
| 15 MHz, 1.5 MHz or 5 kHz bandwidth limiter |
| Number of channels |
2 isolated channels |
4 isolated channels |
| Input impedance |
1 MΩ ± 0.5% , approx. 12 pF |
| Maximum input voltage |
600 V / CAT III (1,000V per Probix) – from 50 to 400 Hz – Probix safety connectors |
| Vertical sensitivity |
16 calibres from 2.5 mV to 200 V/div and up to 156 µV/div in vertical zoom mode (12-bit converter) – Accuracy ± 2% |
| Vertical zoom |
"One Click Winzoom" mode (12-bit converter and direct graphical zoom on screen) – x 16 max |
| Probe factor (non-Probix) |
1 / 10 / 100 / 1,000 or any scaling – definition of measurement unit |
| Horizontal deflection |
| Sweep speed |
35 calibres from 1 ns/div to 200 s/div., accuracy ± [50 ppm + 500 ps] – Roll mode from 100 ms to 200 s/div |
| Horizontal zoom |
"One Click Winzoom" system (direct graphical zoom on screen) x 1 to x 5 or x 100 – storage 100 kpts/channel |
| Triggering |
| Mode |
On all the channels: automatic, triggered, one-shot, auto level 50% |
| Type |
Edge, pulse width (16 ns-20 s), delay (48 ns to 20 s), counting (3 to 16,384 events)
Continuous adjustment of Trigger position |
| Sensitivity |
≤ 1.2 division p-p up to 300 MHz |
| Digital storage |
| Maximum sampling rate |
2.5 GS/s in one-shot mode on each channel (100 GS/s max. in ETS mode) |
| Vertical resolution |
12 bits (vertical resolution 0.025 %) |
| Memory depth |
100 kpts per channel and file viewer in the manager |
| User storage File management |
Internal = 1 GB to store the files: trace, text, configuration, math functions,
System memory: .pdf print files, .png image files
+ high-capacity removable µSD-Card: SD 2 GB, SDHC 4-32 GB and SDXC > 32 GB |
| GLITCH mode |
Duration ≤ 2 ns – 500,000 Min/Max pairs |
| Display modes |
Envelope, vector, accumulation-, averaging (factors 2 to 64) – XY (vector) and Y(f)=FFT |
| Other functions |
| AUTOSET |
Complete in under 5 s, with recognition of the channels – Frequency > 30 Hz |
| FFT analyser & MATH functions |
2,500-point FFT (Lin or Log) with measurement cursors – Functions + , - , x , / and mathematical function editor |
| Cursors |
2 or 3 cursors: simultaneous V and T with AUTO measurement: T1, T2, Dt, 1/Dt, dBV, Ph |
| Automatic measurements |
Simultaneously with waveform, 20 automatic measurements per channel and on the 4 channels simultaneously with scroll |
| MULTIMETER MODE |
| General specifications |
2 or 4 channels – 8,000 cts min/max/frequency/relative – TRMS – Time/date-stamped graphical recording in logger mode |
| AC, DC and AC + DC voltages |
600 mV to 600 VRMS, 800 mV to 800 VDC – VDC accuracy +/- (0.5 % + 25 D) – 200 kHz bandwidth |
| Resistance |
80 Ω to 32 MΩ – accuracy 0.5%R+ 25D – Quick continuity test < 10 ms |
| Other measurements |
Temperature (HX0035 = KTC, HX0036 = Pt100) / Capacitance 5nF to 5mF / Frequency 200 kHz / Diode test 3.3 V |
| Single and three-phase power |
Active, Reactive and Apparent power values plus Power Factor simultaneously with the U & I measurements |
| HARMONIC ANALYSER MODE |
| Multi-channel analysis |
2 or 4 (depending on model), 63 orders, fundamental frequency 40 to 450 Hz in auto or manual mode |
| Simultaneous measurements |
Total Vrms, THD and selected order (% fundamental, phase, frequency, Vrms) |
| HARMONIC ANALYSER MODE |
| Multi-channel analysis |
2 or 4 (depending on model), 63 orders, fundamental frequency 40 to 450 Hz in auto or manual mode |
| Simultaneous measurements |
Total Vrms, THD and selected order (% fundamental, phase, frequency, Vrms) |
| LOGGER MODE |
| Acquisition |
Duration: 20,000 s – Interval: 0.2 s – Files: 100,000 measurements |
| Technical Specifications |
| Configuration memories |
Not limited according to device - variable file sizes |
| Printing |
Network printing via Ethernet/Wifi in .png format |
| PC communication – software link |
Ethernet (100 baseT), WiFi-USB (device, 12 Mbs) – "ScopeNet" application software for PC |
| Software |
PC: Ethernet and USB, ScopeNet (remote control, data recovery, cursors and automatic measurements)
Android tablet – ScopeAdmin Fleet Administration utility |
| Mains power supply |
Li-Ion rechargeable battery (6,900mAH-40 Wh) – Battery life of up to 8 hrs – Adjustable standby mode
Adapter / 2-hour fast charger, universal 98-264 V / 50/60 Hz) |
| Safety / EMC |
Safety as per IEC 61010-2-30, 2010 – 600 V CAT III / 1000 V CAT II – EMC as per EN61326-1, 2010 |
| Mechanical specifications |
292.5 x 210.6 x 66.2 mm – 2.1 kg with batteries – IP54 protection |