| Specifications | 
        
            | Accuracy | ±0.015% of reading (±0.008% optional) | 
        
            | *Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. | 
        
            | Mass and PCU Materials of Construction | 
        
            | Standard weight material | Series 3 non-magnetic austenitic stainless steel | 
        
            | Weight density | 2.7 g/cm³ | 
        
            | Optional fractional weights | Solution heat treated aluminum | 
        
            | Models P3011, P3012, P3013, P3022, P3023, P3025 (V) | Piston Material: Cylinder Material:
 Coefficient of Expansion:
 | Ceramic Martensitic steel
 11 ppm/°C
 | 
        
            | Models P3014, P3015, P3025 (P) P3031, P3032 (P & V) | Piston Material: Cylinder Material:
 Coefficient of Expansion:
 | Tungsten carbide Martensitic steel
 16.5 ppm/°C
 | 
        
            | Model P3016 | Piston Material: Cylinder Material:
 Coefficient of Expansion:
 | Tungsten carbide Tungsten carbide
 11 ppm/°C
 | 
        
            | General | 
        
            | Test port adaptors | 1/8, 1/4, 3/8 and ½ NPT or BSP | 
        
            | Instrument weight | 24 lbs | 
        
            | Instrument size | 17.5 W x 12 D x 8.5 H | 
        
            | Mass set weight (typical) | 65 lbs | 
        
            | Minimum Standard Weight Increments | 
        
            | P3012, P3022 Pressure | 1 inH2O, (5 mbar) | 
        
            | P3013, P3023 Pressure | 2 inH2O, (10 mbar) | 
        
            | P3014, P3015, P3025 Pressure | 1 psi, (0.1bar) | 
        
            | P3016, P3031, P3032 Pressure | 1 psi, (0.1bar) | 
        
            | P3011, P3022, P3023, P3025 Vacuum | 0.2 inHg (10 mbar) | 
        
            | Optional Fractional Weights | 
        
            | P3014, P3015, P3025 Pressure | 0.1 psi (0.01 bar) | 
        
            | Built-in Hand Pumps | 
        
            | Pressure mode | Max pressure 300 psi (20 bar) | 
        
            | Vacuum mode | To 90% vacuum | 
        
            | Operating Fluid for Liquid Lubricated Models | 
        
            | 55-655 | mineral oil (standard) | 
        
            | Krytox | for oxygen-safe applications (optional) | 
        
            | Options | 
        
            | PressCal Software | Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments. | 
        
            | Krytox | As operating fluid for liquid lubricated gas operated models. |