				Fluke Calibration 9173-DW-R-156
                
                
                
 			 
				Catalog #: 3016060
Metrology Well with Built-In Reference, No Insert, 700C, 115V
 
 
                
                    OverviewSpecificationsSummaryCasesInserts 	
                    
				        
                        Product Highlights 
    Best-performing industrial heat sources (accuracy, stability, uniformity) in the worldImmersion depth to 203 mm (8 in)Optional ITS-90 reference input reads PRTs to ±0.006 °CTemperature range from –45 °C to 700 °C Each Unit Includes 
    9173 Metrology WellPower CordRS-232 CableGetting Started GuideUser’s GuideReport of Calibration and calibration label. Accredited ISO17025 CalibrationDIN Connector (-R units only)Tongs (insert removal tool)9930 Interface-it Software and User’s Guide 
                          
    
        
            | Specifications |  
            |  |  |  |  |  |  
            | Range (at 23 °C ambient) | –45 °C to 140 °C (–49 °F to 284 °F)
 | –30 °C to 155 °C (–22 °F to 311 °F)
 | 35 °C to 425 °C (95 °F to 797 °F)
 | 50 °C to 700 °C† (122 °F to 1292 °F)
 |  
            | Display Accuracy | ± 0.1 °C full range | ± 0.1 °C: 35 °C to 100 °C ± 0.15 °C: 100 °C to 225 °C
 ± 0.2 °C: 225 °C to 425 °C
 | ± 0.2 °C: 50 °C to 425 °C ± 0.25 °C: 425 °C to 660 °C
 |  
            | Stability | ± 0.005 °C full range | ± 0.005 °C: 35 °C to 100 °C ± 0.008 °C: 100 °C to 225 °C
 ± 0.01 °C: 225 °C to 425 °C
 | ± 0.005 °C: 50 °C to 100 °C ± 0.01 °C: 100 °C to 425 °C
 ± 0.03 °C: 425 °C to 700 °C
 |  
            | Axial Uniformity (60 mm) | ± 0.1 °C at –45 °C ± 0.04 °C at –35 °C
 ± 0.02 °C at 0 °C
 ± 0.07 °C at 140 °C
 | ± 0.025 °C at –30 °C ± 0.02 °C at 0 °C
 ± 0.07 °C at 155 °C
 | ± 0.05 °C: 35 °C to 100 °C ± 0.1 °C: 100 °C to 225 °C
 ± 0.2 °C: 225 °C to 425 °C
 | ± 0.1 °C: 50 °C to 100 °C ± 0.25 °C: 100 °C to 425 °C
 ± 0.4 °C: 425 °C to 700 °C
 |  
            | Radial Uniformity | ± 0.01 °C full range | ± 0.01 °C: 35 °C to 100 °C ± 0.02 °C: 100 °C to 225 °C
 ± 0.025 °C: 225 °C to 425 °C
 | ± 0.01 °C: 50 °C to 100 °C ± 0.025 °C: 100 °C to 425 °C
 ± 0.04 °C: 425 °C to 700 °C
 |  
            | Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes) | ± 0.02 °C at –45 °C ± 0.005 °C at –35 °C
 ± 0.01 °C at 140 °C
 | ± 0.005 °C at –30 °C ± 0.005 °C at 0 °C
 ± 0.01 °C at 155 °C
 | ± 0.01 °C full range | ± 0.02 °C at 425 °C ± 0.04 °C at 700 °C
 |  
            | Hysteresis | 0.025 °C | 0.04 °C | 0.07 °C |  
            | Well Depth | 160 mm (6.3 in) | 203 mm (8 in) |  
            | Resolution | 0.001 °C |  
            | Display | LCD, °C or °F, user-selectable |  
            | Key Pad | Ten key with decimal and +/- button. Function keys, menu key, and °C / °F key. |  
            | Cooling Time | 44 min: 23 °C to –45 °C 19 min: 23 °C to –30 °C
 19 min: 140 °C to 23 °C
 | 30 min: 23 °C to –30 °C 25 min: 155 °C to 23 °C
 | 220 min: 425 °C to 35 °C 100 min: 425 °C to 100 °C
 | 235 min: 700 °C to 50 °C 153 min: 700 °C to 100 °C
 |  
            | Heating Time | 32 min: 23 °C to 140 °C 45 min: –45 °C to 140 °C
 | 44 min: 23 °C to 155 °C 56 min: –30 °C to 155 °C
 | 27 min: 35 °C to 425 °C | 46 min: 50 °C to 700 °C |  
            | Size (HxWxD) | 366 x 203 x 323 mm (14.4 x 8 x 12.7 in) |  
            | Weight | 14.2 kg (31.5 lb) | 15 kg (33 lb) | 13.2 kg (29 lb) | 15 kg (33 lb) |  
            | Power | 115 V ac (± 10 %), or 230 V ac (± 10 %), 50/60 Hz, 550 W
 | 115 V ac (± 10 %), or 230 V ac (± 10 %), 50/60 Hz, 1025 W
 |  
            | Computer Interface | RS-232 Interface with 9930 Interface-it control software included |  
            | Traceable Calibration (NIST) | Data at –45 °C, 0 °C, 50 °C, 100 °C, and 140 °C | Data at –30 °C, 0 °C, 50 °C, 100 °C, and 155 °C | Data at 100 °C, 150 °C, 250 °C, 350 °C, and 425 °C | Data at 100 °C, 200 °C, 350 °C, 500 °C, and 660 °C |  
            | †Calibrated to 660 °C; reference thermometer recommended at higher temperatures. |  
            | Specifications | Built-in Reference Input |  
            | Temperature Range | –200 °C to 962 °C (–328 °F to 1764 °F) |  
            | Resistance Range | 0 Ω to 400 Ω, auto-ranging |  
            | Characterizations | ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 Callendar-Van Dusen (CVD): R0, a, b, d |  
            | Resistance Accuracy | 0 Ω to 20 Ω: 0.0005 W 20 Ω to 400 Ω: 25 ppm
 |  
            | Temperature Accuracy (does not include probe uncertainty) | 10 Ω PRTs: ± 0.013 °C at 0 °C
 ± 0.014 °C at 155 °C
 ± 0.019 °C at 425 °C
 ± 0.028 °C at 700 °C
 | 25 Ω and 100 Ω PRTs: ± 0.005 °C at –100 °C
 ± 0.007 °C at 0 °C
 ± 0.011 °C at 155 °C
 ± 0.013 °C at 225 °C
 ± 0.019 °C at 425 °C
 ± 0.027 °C at 661 °C
 |  
            | Resistance Resolution | 0 Ω to 20 Ω: 0.0001 Ω 20 Ω to 400 Ω: 0.001 Ω
 |  
            | Measurement Period | 1 second |  
            | Probe Connection | 4-wire with shield, 5-pin DIN connector |  
            | Calibration | NVLAP accredited (built-in reference input only), NIST-traceable calibration provided |    
    
        
            | Calibration Baths |  
            | Type | Dry |  
            | Min Temp (Caiibration Baths) | 50 C |  
            | Max Temp (Calibration Baths) | 700 C |  
            | Accuracy (Calibration Baths) | 0.2 C |  
            | Voltage (Calibration Baths) | 115 VAC |  
            | Special Features (Calibration Baths) | Built in Readout for Probe, Inserts Available (Dry Baths), PC Software |  
         
            | Calibration Included | Factory Calibration |  
            | Power Supply Voltage | 120V 50/60 Hz |  
            | Interfaces I/O | RS-232 |  
            | ECCN Number | EAR99 |  
            | Country of Origin | United States |  |  
				 |  
     |  |