                    OverviewSpecificationsSummaryCasesAccessories 	
                    
				        
                        Product Highlights
    Three sensor types
    
        PRTs: –200 °C to 1000 °CThermocouples –200 °C to 2315 °CPrecision thermistors: –50 °C to 150 °CAccuracy
    
        PRTs: up to ±0.011 °CThermocouples : ±0.24 °C for J,K,L,MPrecision thermistors: ±0.002 °CFast mode
    
        PRTs: 0.45 seconds per sampleThermocouples: 0.3 second per samplePrecision thermistors: 0.3 seconds per sampleGraphical display
    
        128x64 backlit LCD graphic displayPlot and scale trends in real timeSimultaneous dual channel reading Each Unit Includes 
    1523-156 Reference Thermometer Readout with Accredited Calibration(1) 5615-12-P Platinum Resistance Thermometer -200 to 420°C (-328 to 788°F) with Accredited CalibrationUniversal TC INFO-CON ConnectorTPAK1523-CaseUsers GuideCD-ROM (Contains Technical Manual)12 V DC Universal Power SupplyRS-232 Cable9940 I/O ToolKit Software 
                          
    
        
            | General Specifications |  
            |  | 1523 | 1524 |  
            | Input Channels | 1 | 2 |  
            | Logging | 25 readings with statistics | 25 readings with statistics 15,000 time and date stamped
 |  
            | Sample interval (normal): | 1 second | 1 second (simultaneous measurement) |  
            | Typical sample interval (fast mode)*: | 0.3 seconds | 0.3 seconds |  
            | Sensor types | PRTs, RTDs, Thermistors, and Thermocouples |  
            | Thermocouple types | C,E,J,K,L,M,N,T,U,B,R,S |  
            | Operating temperature | -10 °C to 60 °C (Best accuracy 13 °C to 33 °C) |  
            | Power requirements | 3 AA alkaline batteries |  
            | Safety | EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 |  
            | Size (HxWxL) | 96 mm x 200 mm x 47 mm (3.75 in x 7.9 in x 1.86 in)
 |  
            | Weight | 0.65 kg (1.4 lb) |  
            | Environmental conditions for best accuracy: 13 °C to 33 °C |  
            | Millivolt range and accuracy | -10 mV to 75 mV ± (0.005 % + 5 V)
 |  
            | Resistance range and accuracy | 0 Ω to 400 Ω ± (0.004 % + 0.002 Ω) 200 Ω to 50 kΩ ± (0.01 % + 0.5 Ω)
 50 kΩ to 500 kΩ ± (0.03 %)
 |  
            | Temperature coefficient, voltage ( -10 °C to 13 °C, + 33 °C to 60 °C) | Temperature coefficient ( -10 °C to 13 °C , +33 °C to 60 °C):
 ± (0.001 %/°C + 1 V/°C)
 |  
            | Temperature Coefficient, Resistance ( -10 °C to 13 °C, + 33 °C to 60 °C) | 0.0008 %/ °C + 0.0004 Ω (0 Ω to 400 Ω) 0.002 %/ °C + 0.1 Ω (0 Ω to 50 kΩ)
 0.06 %/ °C + 0.1 Ω (50 kΩ to 500 kΩ)
 |  
            | Excitation current, resistance | 1 mA (0 Ω to 400 Ω) 10 µA (0 Ω to 50 kΩ)
 2 µA (50 kΩ to 500 kΩ)
 |  
            | *See technical manual for sample interval details by sensor type and number of inputs. |  
    
        
            | 1523/24 Accuracies with selected probes (± °C) |  
            | T(°C) | 5616-12 | 5615-6 | 5627A-12 | 5610-9 |  
            | -200 | 0.014 | 0.025 | 0.027 | n/a |  
            | 0 | 0.021 | 0.021 | 0.049 | 0.009 |  
            | 100 | 0.027 | 0.028 | 0.065 | 0.030 |  
            | 300 | 0.040 | 0.043 | 0.103 | n/a |  
            | 420 | 0.050 | n/a | 0.130 | n/a |  
            | Includes readout accuracy, probe calibration, and probe drift |  
    
        
            | Temperature, thermocouples |  
            | Type | Range | Measure accuracies |  
            | B | 600 °C to 800 °C 800 °C to 1000 °C
 1000 °C to 1800 °C
 | 0.85 °C 0.68 °C
 0.57 °C
 |  
            | C | 100 °C to 550 °C 550 °C to 2300 °C
 | 0.32 °C 0.71 °C
 |  
            | E | -200 °C to 0 °C 0 °C to 950 °C
 | 0.52 °C 0.22 °C
 |  
            | J | -200 °C to 0 °C 0 °C to 1200 °C
 | 0.52 °C 0.23 °C
 |  
            | K | -200 °C to 0 °C 0 °C to 1370 °C
 | 0.61 °C 0.24 °C
 |  
            | L | -200 °C to 0 °C 0 °C to 900 °C
 | 0.36 °C 0.23 °C
 |  
            | M | -20 °C to 0 °C 0 °C to 400 °C
 400 °C to 1400 °C
 | 0.26 °C 0.25 °C
 0.22 °C
 |  
            | N | -200 °C to 0 °C 0 °C to 1300 °C
 | 0.72 °C 0.28 °C
 |  
            | R | -20 °C to 0 °C 0 °C to 500 °C
 500 °C to 1750 °C
 | 1.09 °C 0.97 °C
 0.49 °C
 |  
            | S | -20 °C to 0 °C 0 °C to 500 °C
 500 °C to 1750 °C
 | 1.05 °C 0.95 °C
 0.56 °C
 |  
            | T | -200 °C to 0 °C 0 °C to 400 °C
 | 0.60 °C 0.25 °C
 |  
            | U | -200 °C to 0 °C 0 °C to 400 °C
 | 0.54 °C 0.24 °C
 |  
            | Resolution: |  
            | B, C, E, J, K, L, M, N, R, S, T, U: | 0.10 °C |  
            | Note 1: Accuracies are based on internal reference junction compensation. Refer to technical manual for equivalent accuracies with an external reference junction. |    
    
        
         
            | Calibration Included | NIST Calibration |  
            | Safety Approval | CSA |  
            | Data Logging | Yes |  
            | Interfaces I/O | RS-232 |  
            | Product Weight | 1.4 LBS |  
            | Product Height | 47 MM (1.85 IN) |  
            | Product Length | 96 MM (3.78 IN) |  
            | Product Width | 200 MM (7.87 IN) |  
            | ECCN Number | EAR99 |  
            | Country of Origin | United States |  |  
				 |  
     |  |