				Fluke Calibration 7109A-P-156
                
                
                
 			 
				Catalog #: 4757585
Portable Calibration Bath, -25 C TO 140 C, W/PROC ELEC, 115 VAC
 
 
                
                    OverviewSpecificationsSummaryAccessories 	
                    
				        
                        Product Highlights
    Portable Bath, -25 C to 140 C, with Process Electronics, 115 VACCalibrate up to four tri-clamp sanitary sensors at the same timeWide temperature ranges cover most process applications:6109A: 35 °C to 250 °C7109A: –25 °C to 140 °CExcellent display accuracy of ± 0.1 °C provides 4:1 test uncertainty ratio (TUR) for critical applicationsEasy to transport up stairs and across catwalksStainless steel casing withstands harsh sterilizing chemicals and is rust proofEasy to use and maintainGlobal Fluke support and service Each Unit Includes 
    7109A-P-156USB cable        Probe access coverTransport tank cover Printed safety information  Product CD (includes Operators Manual, USB Driver, Software distribution license, Source code for public domain software)NVLAP accredited calibration certificateClamp on FerriteDIN ConnectorTest Lead Kit 
                        
    
        
            | General specifications |  
            | Size | Height: 382 mm (15 in) Width: 242 mm (9.5 in) Depth: 400 mm (15.7 in) |  
            | Weight | 6109A: 16 kg (35 lb) 7109A: 20 kg (44 lb) |  
            | Fluid volume | 2.5 liters (0.66 gallons) |  
            | Fluid working area | 75 mm x 75 mm (3 in x 3 in) |  
            | Maximum fluid depth | 154 mm (6.1 in) |  
            | Factory calibration | Traceable NVLAP-accredited calibration included |  
            | Temperature Specifications |  
            |  | 6109A | 7109A |  
            | Range (at 25 °C ambient temperature) | 35 °C to 250 °C | -25 °C to 140 °C (–15 °C cover off) |  
            | Accuracy (maximum permissible error) | 0.1 °C | 0.1 °C |  
            | Display resolution | 0.01°, 0.001° | 0.01°, 0.001° |  
            | Stability | 0.015 °C | 0.01 °C |  
            | Typical uniformity | 0.03 °C up to 200 °C 0.04 °C above 200 °C
 | 0.02 °C |  
            | Repeatability | 0.04 °C | 0.04 °C |  
            | Typical heating time | 35 °C to 100 °C: 25 minutes 100 °C to 250 °C: 45 minutes
 | -25 °C to 25 °C: 35 minutes 25 °C to 140 °C: 55 minutes
 |  
            | Typical cooling time | 250 °C to 100 °C: 35 minutes 100 °C to 35 °C: 55 minutes
 | 140 °C to 25 °C: 45 minutes 25 °C to –25 °C: 75 minutes
 |  
            | Typical settling time | 15 minutes | 10 minutes |  
            | Process Input Module Specifications (-P Models) |  
            | Function | Range | Accuracy (Maximum Permissible Error) |  
            | Reference input resistance | 0Ω to 42Ω 42 Ω to 400 Ω
 | 0.0025 Ω 0.006 %
 |  
            | Reference input temperature | -25 °C to 250 °C | 0.007 % + 0.015 °C |  
            | Resistance sensing current | 1 mA | 8% |  
            | DUT 4-wire resistance | 0 Ω to 31 Ω 31 Ω to 400 Ω
 | 0.0025 Ω 0.008 %
 |  
            | DUT 3-wire resistance accuracy | 0 Ω to 400 Ω | 0.12 Ω |  
            | Thermocouple mV | -10 mV to 100 mV | 0.025 % + 0.01 mV |  
            | Reference junction temperature | 0 °C to 40 °C | 0.35 °C |  
            | Thermocouple temperature | -25 °C to 250 °C | J: 0.44 °C   K: 0.49 °C   T: 0.53 °C E: 0.44 °C   N: 0.57 °C   M: 0.48 °C
 L: 0.42 °C   U: 0.48 °C
 |  
            | 0 °C to 250 °C | R: 1.92 °C   S: 1.88 °C   C: 0.84 °C D: 1.12 °C   G/W: 3.34 °C
 |  
            | mA range | 0 mA to 22 mA | 0.02 % + 0.002 mA |  
            | Loop power | 24 V dc | ± 6 V |    
    
        
            | Calibration Baths |  
            | Type | Liquid |  
            | Min Temp (Caiibration Baths) | -25 C |  
            | Max Temp (Calibration Baths) | 140 C |  
            | Accuracy (Calibration Baths) | 0.1 C |  
            | Voltage (Calibration Baths) | 115 VAC |  
            | Special Features (Calibration Baths) | Built in Readout for Probe, PC Software, Programmable Tests |  
         
            | Calibration Included | Factory Calibration |  
            | Power Supply Voltage | 120V 50/60 Hz |  
            | Interfaces I/O | RS-232, USB |  
            | Product Weight | 44 LBS |  
            | ECCN Number | EAR99 |  
            | Country of Origin | United States |  |  
				 |  
     |  |