| Electrical Specifications | 
        
            | Maximum voltage to earth ground: | 1000 V | 
        
            | Battery type: | AA alkaline IEC/EN LR6 (x4) | 
        
            | Battery life: | More than 15 hours* | 
        
            | Frequency range: | 40 Hz to 1 kHz | 
        
            | Ingress protection: | IEC/EN 60529: IP30 with jaw closed | 
        
            | LCD: | Digital reading: 9999 counts Refresh rate: 4 per second
 | 
        
            | Operating temperature: | -10 °C to + 50 °C | 
        
            | Storage temperature: | -20 °C to + 60 °C | 
        
            | Operating humidity: | Non condensing (<10 °C) (< 50 °F) ≤90% RH (at 10 °C to 30 °C [50 °F to 86 °F])
 ≤75% RH (at 30 °C to 40 °C [86 °F to 104 °F])
 ≤45% RH (at 40 °C to 50 °C [104 °F to 122 °F])
 (Non condensing)
 | 
        
            | Operating altitude: | 2000 m (6561 feet) | 
        
            | Storage altitude: | 12 000 m (39,370 feet) | 
        
            | Reference temperature: | 23 °C ±5 °C (73 °F ±9 °F) | 
        
            | Temperature coefficient: | 0.15% x (specified accuracy)/ °C (<18 °C or >28 °C [<64.4 °F or >82.4 °F]) | 
        
            | Overload indication: | OL | 
        
            | Data logging capacity: | Minimum of 32,760 measurements | 
        
            | Data logging interval: | 1 second to 59 minutes and 59 seconds | 
        
            | Display reading specification with standard loop resistance | 
        
            | 
                
                    
                        | Input (Ω) | Minimum | Maximum |  
                        | 0.474 | 0.417 | 0.531 |  
                        | 0.5 | 0.443 | 0.558 |  
                        | 10 | 9.55 | 10.45 |  
                        | 100 | 96 | 104 |  | 
        
            | Safety | 
        
            | General: | IEC/EN 61010-1: Pollution Degree 2 IEC/EN 61557-1 | 
        
            | Measurement: | IEC/EN 61010-2-032: CAT IV 600 V / CAT III 1000 V | 
        
            | Current clamp for leakage current measurements: | IEC/EN 61557-13: Class 2, ≤30 A/m | 
        
            | Resistance to earth: | IEC/EN 61557-5 | 
        
            | Effectiveness of thepProtective measures: | IEC/EN 61557-16 | 
        
            | Electromagnetic compatibility (EMC) | International: IEC/EN 61326-1: Portable Electromagnetic Environment CISPR 11: Group 1, Class B, IEC/EN 61326-2-2
 | 
        
            | Korea (KCC): Class A equipment (Industrial Broadcast & Communications Equipment) USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103
 | 
        
            | Wireless Radio | 
        
            | Frequency range: | 2412 MHz to 2462 MHz | 
        
            | Output power: | <10 mW | 
        
            | Radio frequency certification: | FCC ID:T68-FBLE IC:6627A-FBLE | 
        
            | General Specifications | 
        
            | Conductor size: | 40 mm (1.57 in) approximately | 
        
            | Dimensions (L x W x H): | 283 x 105 x 48 mm (11.1 x 4.1 x 1.9 in) | 
        
            | Weight: | 880g (31 oz) | 
        
            | Warranty: | One year | 
        
            | Earth Ground Loop Resistance | 
        
            | 
                
                    1 Loop resistance with no inductance, conductor centered and perpendicular to jaw
                        | Range | Accuracy1 ± (% of reading + Ω) |  
                        | 0.025 Ω to 0.249 Ω: | 1.5% + 0.02 Ω |  
                        | 0.250 Ω to 0.999 Ω: | 1.5% + 0.05 Ω |  
                        | 1.000 Ω to 9.999 Ω: | 1.5% + 0.10 Ω |  
                        | 10.00 Ω to 49.99 Ω: | 1.5% + 0.30 Ω |  
                        | 50.00 Ω to 99.99 Ω: | 1.5% + 0.50 Ω |  
                        | 100.0 Ω to 199.9 Ω: | 3.0% + 1.0 Ω |  
                        | 200.0 Ω to 399.9 Ω: | 5.0% + 5.0 Ω |  
                        | 400 Ω to 599 Ω: | 10.0% + 10 Ω |  
                        | 600 Ω to 1500 Ω: | 20.00% |  | 
        
            | 
                
                    1 Applies to signal frequency
                        | Range | Resolution | Accuracy1 ± (% of reading + mA) |  
                        | 0.200 mA to 3.999 mA | 1 μA | 2.0% + 0.05 mA |  
                        | 4.00 mA to 39.99 mA | 10 μA | 2.0% + 0.03 mA |  
                        | 40.0 mA to 399.9 mA | 100 μA | 2.0% + 0.3 mA |  
                        | 0.400 A to 3.999 A | 1 mA | 2.0% + 3 mA |  
                        | 4.00 A to 39.99 A | 10 mA | 2.0% + 30 mA |  • 40 Hz to 1 kHz with filter set to OFF