				Hioki 8969
                
                
                
 			 
				2ch, Distortion Measurement for Strain Gauge Converter
			 
 
				
                Product HighlightsMeasure strain waveforms with 16-bit resolution
Directly connect to the strain gauge converter (bridge voltage: DC only) 
16-bit A/D high-resolution 
Isolated input, up to 33V AC, or 70V DC from the earth 
Meet application demands in the civil engineering, bridge building and machine tool and automotive design fields
Use with the MEMORY HiCORDER 8847
Accuracy at 23 ±5 °C/73 ±9 °F, 20 to 80 % rh after 30 minutes of warm-up time and auto-balance; accuracy/product guaranteed for 1 year)
Measurement functions Number of channels: 2, for distortion measurement (electronic auto-balancing, balance adjustment range within ±10000µe) 
Input connectors Weidmuller SL 3.5/7/90G (via Conversion Cable 9769, TAJIMI PRC03-12A10-7M10.5)
Max. rated voltage to earth: 33Vrms or 70V DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) 
Suitable transducer Strain gauge converter, Bridge impedance: 120? to 1k?, Bridge voltage: 2V ±0.05V, Gauge rate: 2.0 
Measurement range 20µe to 1000µe/div, 6ranges, full scale: 20division, Low-pass filter: 5/10/100Hz, 1kHz
Measurement resolution 1/1250 of measurement range (using 16-bit A/D conversion and when installed in the 8847) 
Highest sampling rate 200kS/s (simultaneous sampling across 2 channels) 
Measurement accuracy ±(0.5% of full scale +4µe) (at 5Hz filter ON, After auto-balancing) 
Frequency characteristics DC to 20kHz +1/-3dB 
Dimensions, mass 106mm (4.17in) W × 19.8mm (0.78in) H × 196.5mm (7.74in) D, 220g (7.8oz)
 
   Each Unit IncludesConversion cable (cable length 50cm/1.64ft)
 
   |  
					
				 
					
				 
				 |  
     |  |