				Hioki 8946
                
                
                
 			 
				4ch Analog Unit for 8826, 8860/ 61-50
			 
 
				
                Product HighlightsThis UNIT is equipped with circuits for four independently usable input channels. Simply slot the 8946 into the MEMORY HiCORDER for use
Measurement ranges 10 mV to 2 V/division, 8 ranges, the measurement resolution is 1/80 of range
* When attached to recorder where full-scale (f.s.) = 20 divisions
20 mV to 5 V/division, 8 ranges, the measurement resolution is 1/160 of range
* When attached to recorder where full-scale (f.s.) = 10 divisions
Low-pass filter: 5/500/5 k/50 kHz 
Maximum sampling rate 1 MS/s (simultaneous sampling of four channels) 
Accuracy DC amplitude: ±0.5 % f.s.   Zero-position: ±0.15 % f.s. 
Zero-position -50 % to 150 %, 1 % step * With zero-adjustment function 
Frequency characteristics DC to 100 kHz ±3 dB 
Max. allowable input 30 V rms or 60 V DC   (upper voltage which when applied to between input pins does not damage them) 
Max. rated voltage to earth 30 V rms or 60 V DC   (upper voltage which when applied to input channel casing or between input channels does not damage them)
 
   |  
					
				 
					
				 
				 |  
     |  |