				Hioki 8938
                
                
                
 			 
				FFT Analog Unit
			 
 
				
                Product HighlightsSimply slot the 8938 into the MEMORY HiCORDER for use
Input Number of channels: 2, Connector: Insulated BNC
* Input isolated from output, inter-channel isolation 
Measurement ranges 5 mV to 20 V/division, 12 ranges, the measurement resolution is 1/80 of range
AC voltage for possible measurement/display using the memory function: 280 V rms
* When attached to recorder where full-scale (f.s.) = 20 divisions
10 mV to 50 V/division, 12 ranges, the measurement resolution is 1/160 of range
AC voltage for possible measurement/display using the memory function: 280 V rms
* When attached to recorder where full-scale (f.s.) = 10 divisions
Low-pass filter: 5/500/5 k/100 kHz 
Anti-aliasing filter Cutoff frequency 20/40/80/200/400/800/2 k/4 k/8 k/20 k/40 kHz auto-select (linked to frequency range) 
Maximum sampling rate 1 MS/s (simultaneous sampling of two channels) 
Accuracy DC amplitude: ±0.4 %f.s.   Zero-position: ±0.1 %f.s. 
Zero-position -50 % to 150 %, 1 % step * With zero-adjustment function 
Frequency characteristics DC to 400 kHz ±3 dB   with AC coupling: 7 Hz to 400 kHz ±3 dB 
Max. allowable input 400 V DC   (upper voltage which when applied to between input pins does not damage them) 
Max. rated voltage to earth 370 V AC, DC   (upper voltage which when applied to input channel casing or between input channels does not damage them) 
 
   |  
					
				 
					
				 
				 |  
     |  |