				Fluke Calibration 5611T-J
                
                
                
 			 
				Catalog #: 2740050
Teflon Reference, Bare Thermistor Probe, Banana Plugs
 
 
                
                    OverviewSpecificationsSummaryCases 	
                    
				        
                        Product Highlights
    Banana plug connector tipsProbe tip: 0.12 inch diameter and 1.1 inch probe tip lengthShort-term accuracy to +/- 0.01 °C; one-year drift < +/-0.01 °CAccredited NVLAP calibration optionalPTFE encapsulated thermistor Each Unit Includes 
    Coefficient SheetCertification Sheets 
                          
    
        
            | Specifications |  
            | Parameter | Value |  
            | Maximum temperature | 100 °C |  
            | Minimum temperature | 0 °C |  
            | Nominal resistance 25 °C | 10,000 W ± 0.01 |  
            | Temperature coefficient | Negative (NTC) |  
            | Length of sensor | 13 mm (0.51 in) |  
            | Length of sheath | 2.79 cm (1.10 in) |  
            | Diameter of sheath | 3.0 mm (0.12 in) |  
            | Sheath material | Teflon |  
            | Immersible length of instrument | ~ 6 m (20 ft) |  
            | Short term stability maximum temp (no less than three thermal cycles) | <6 mK |  
            | Long term stability maximum temp (no less than 100 hrs @ max temp) | <6 mK |  
            | Hysteresis (if different from short term stability) | <1 mK |  
            | Self heating (DC in 25°C bath) | 4 mW/°C typical |  
            | Self heating (DC in room temperature air) | .014 mW/°C typical |  
            | Time constant (room temperature to 70°C bath) | 4 seconds typical |  
            | Time constant (70°C bath to room temperature) | 30 seconds typical |  
            | Susceptibility to vibration (lo med hi) | Low |  
            | Lead wire cable type | 4 wire, 30 AWG, 25/44, clear FEP Teflon insulation |  
            | Lead wire outside diameter | 1.75 mm (0.069 in) |  
            | Lead wire length | 6.1 m (20 ft) |  
            | Standard calibration | NIST traceable calibration, R vs. T table with 0.1 °C increments, interpolation equation furnished |  
            | Calibration uncertainty | Table and equation are accurate to ± 0.01 °C |  
            | Optional calibration | Certificate of NVLAP-accredited calibration, Data supplied at 6 points over 100 °C span, complies with NCSL/ISO/IEC 17025:2000 and ANSI/ NCSL Z540-1-1994 |  
            | Termination | Specify when ordering |    
    
        
         
            | Country of Origin | United States |  |  
				 |  
     |  |