				Fluke Calibration 9101-156
                
                
                
 			 
				Zero - Point Dry-Well NIST 0°C (32°F)
			 
 
				
                Product HighlightsHart Model # 9101 
Bath-quality stability in a portable ice point referance 
Easy recalibration for long term reliability 
Ready light frees user's time and attention 
Solid state cooling technology 
Temperature range: 0°C (32°F) 
Stability: ±0.005°C 
Total Instrument Error: ±0.02°C , typical; ±0.05°C max. (18-25°C ambient) 
Stabilization Time: Approx 30 minutes (the ready lamp indicates stable control at 0°C) 
Temperature Coefficient: ±0.005°C/°C 
Size: 12.25"H x 8.5"W x 5.75"D (311 x 216 x 146 mm) 
Power: 115 VAC (±10%), 1A or 230 VAC (±10%), 0.5A, specify, 50/60Hz, 125W 
Well Dimensions: 2 wells 0.25" dia. x 6"D (6.4 x 152mm), 1 well 0.28" dia. x 6"D (7.152 mm) 
Includes one set of telescoping inserts to provide various small diameters 
Weight: 12lb (5.4kg) 
NIST Traceable Calibration: Data at 0°C
 
   |  
					
				 
					
				 
				 |  
     |  |