				Fluke Calibration P3014-PSI-P
                
                
                
 			 
				Catalog #: 4184411
Air Deadweight Tester Single PCU, 150 PSI, W/ PUMP
 
 
                
                    OverviewSpecificationsSummarysoftwareAccessories 	
                    
				        
                        Product Highlights
    Air Operated – Single PCUPressure Range 3 to 150 psiAccuracy better than 0.015 % of reading. (Increased accuracy option of 0.008 %)Gas operated with optional hand pumpsDual vacuum/pressure models availablePsi, bar, kgf/cm 2 , kPa and MPa ranges availablePiston/cylinder design provides stability and repeatabilityBuilt-in pneumatic hand pumps for pressure and vacuumHigh quality needle valves provide optimum controlMounted spirit level and adjustable feetTest station design with O-rings eliminates the need for PTFE tape or wrenchesImproved case with spring-loaded latches for hoodSturdy weight box with hinged lid and side handles for easy transport 
                        
    
        
            | Specifications |  
            | Accuracy | ±0.015% of reading (±0.008% optional) |  
            | *Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. |  
            | Mass and PCU Materials of Construction |  
            | Standard weight material | Series 3 non-magnetic austenitic stainless steel |  
            | Weight density | 2.7 g/cm³ |  
            | Optional fractional weights | Solution heat treated aluminum |  
            | Models P3011, P3012, P3013, P3022, P3023, P3025 (V) | Piston Material: Cylinder Material:
 Coefficient of Expansion:
 | Ceramic Martensitic steel
 11 ppm/°C
 |  
            | Models P3014, P3015, P3025 (P), P3031, P3032 (P & V) | Piston Material: Cylinder Material:
 Coefficient of Expansion:
 | Tungsten carbide Martensitic steel
 16.5 ppm/°C
 |  
            | Model | Piston Material: Cylinder Material:
 Coefficient of Expansion:
 | Tungsten carbide Tungsten carbide
 11 ppm/°C
 |  
            | General |  
            | Test port adaptors | 1/8, 1/4, 3/8 and ½ NPT or BSP |  
            | Instrument weight | 24 lbs |  
            | Instrument size | 17.5 W x 12 D x 8.5 H |  
            | Mass set weight (typical) | 65 lbs |  
            | Minimum Standard Weight Increments |  
            | P3012, P3022 Pressure | 1 inH2O, (5 mbar) |  
            | P3013, P3023 Pressure | 2 inH2O, (10 mbar) |  
            | P3014, P3015, P3025 Pressure
 | 1 psi, (0.1bar) |  
            | P3016, P3031, P3032 Pressure
 | 1 psi, (0.1bar) |  
            | P3011, P3022, P3023, P3025 Vacuum | 0.2 inHg (10 mbar) |  
            | Optional Fractional Weights |  
            | P3011, P3022, P3023, P3025Pressure
 | 0.1 psi (0.01 bar) |  
            | Built-in Hand Pumps |  
            | Pressure mode | Max pressure 300 psi (20 bar) |  
            | Vacuum mode | To 90% vacuum |  
            | Operating Fluid for Liquid Lubricated Models |  
            | 55-655 | mineral oil (standard) |  
            | Krytox | for oxygen-safe applications (optional) |  
            | Options |  
            | PressCal Software | Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments. |  
            | Krytox | As operating fluid for liquid lubricated gas operated models. |  
    
        
         
            | ECCN Number | EAR99 |  
            | Country of Origin | United States |    |  
				 |  
     |  |